Published June 1995 | Version v1
Journal article

A study on the measurement of effective energy of scattering X-rays

  • 1. Kyoto Coll. of Medical Technology (Japan)

Description

Only a few studies have been reported on the measurement and evaluation of the effective energy of scattering X-rays using an ionization chamber. The reason for this is due to the difficulty in accurately measuring attenuation curve in scattering X-rays lacking any directional properties. We could come up with a new method for calculating the effective energy of scattering X-rays by utilizing their spectra data. First, for analysing the accuracy of our calculation method with using primary X-rays, a comparison was made of calculated values of the effective energy obtained by our calculation method with the measurement values obtained using an ionization chamber. The results gave the calculated values agreeing with the measurement values within a maximum error of 2%, and this method was very helpful in measuring the effective energy of the scattering X-rays. Consequently, this method was capable of measuring the effective energy of scattering X-rays in the following parameters: X-ray tube voltage, scattering angle and size of scatterer. In conclusion, it is considered that our method could solve the present difficulty regarding the measurement of effective energy of the scattering X-rays, and provided a useful procedure concerning the study of radiation protection. (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
51
Journal Issue
6
Journal Page Range
p. 695-701.
ISSN
0369-4305
CODEN
NIPHAP