Published June 14, 1986 | Version v1
Journal article

A liquid metal ion source analysis system

Creators

  • 1. UKAEA Atomic Energy Research Establishment, Harwell. Materials Development Div.

Description

An analysis system for angular and mass resolved energy distribution measurements of liquid metal ion source beams has been constructed. The energy analyser has been calibrated, and preliminary on-axis energy distribution measurements of a gallium source operating between 0.26 and 30.0 μA have been made. These results closely agree with measurements reported by other workers, confirming the unusual FWHM behaviour of gallium sources below approx. 2 μA. (author)

Additional details

Publishing Information

Journal Title
J. Phys., D (London). Appl. Phys.
Journal Volume
19
Journal Issue
6
Series
J. Phys., D (London). Appl. Phys.
Journal Page Range
L115-L119
ISSN
0022-3727
CODEN
JPAPB

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
17063516
Subject category
S74: ATOMIC AND MOLECULAR PHYSICS;
Descriptors DEI
BEAM CURRENTS; CALIBRATION; COMPARATIVE EVALUATIONS; ENERGY SPECTRA; GALLIUM; ION BEAMS; ION SOURCES; LIQUID METALS; MEASURING INSTRUMENTS; SPECIFICATIONS
Descriptors DEC
BEAMS; CURRENTS; ELEMENTS; EVALUATION; FLUIDS; LIQUIDS; METALS; SPECTRA