Study on opening pressure of hydrocephalus shunt under siphon
- 1. School of Mechanical Engineering, University of Jinan, Jinan, Shandong (China)
- 2. Shandong Quality Inspection Center for Medical Devices, Jinan, Shandong (China)
Description
The opening pressure of the hydrocephalus shunt with the siphon device was studied and analyzed in this paper. The influence of fluid velocity and siphon height on the opening pressure of hydrocephalus shunt is analyzed. According to the national medical industry standards, the hydrocephalus shunt must be opened by the liquid column detection method. In this paper, computational fluid dynamics was used to simulate the liquid column pressure, and the interaction analysis of the two factors of flow rate and siphon height was carried out. Finally, the influence of flow velocity and siphon height on the opening pressure of hydrocephalus shunt was obtained in this paper, and the relevant formula of the influence of these two factors on the opening pressure was obtained. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1742-6596/1885/4/042063Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 1885
- Journal Issue
- 4
- Journal Page Range
- [6 p.]
- ISSN
- 1742-6596
Conference
- Title
- 7. International Conference on Manufacturing Technology and Applied Materials
- Acronym
- ICAMMT 2021
- Dates
- 26-28 Mar 2021
- Place
- Guilin (China)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 53082167
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; S42: ENGINEERING;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- DETECTION; FLOW RATE; FLUID MECHANICS; LIQUIDS; MALFORMATIONS
- Descriptors DEC
- FLUIDS; MECHANICS; PATHOLOGICAL CHANGES