Published August 1976 | Version v1
Journal article

Radiation incidents with a cobalt-60 teletherapy unit

  • 1. Wisconsin Univ., Madison (USA). Dept. of Radiology

Description

Three radiation incidents associated with source drive failures on one particular 60Co teletherapy unit, which took place during a single year's operation, are discussed. In each case the extra treatment time of the patient, the exposure to the technician, the cause of the breakdown and the subsequent action taken are detailed. Recommendations for users of 60Co teletherapy units to minimise the results of the malfunctioning of such a system are given. (U.K.)

Additional details

Publishing Information

Journal Title
Health Phys.
Journal Volume
31
Journal Issue
2
Series
Health Phys.
Journal Page Range
163-165