Published 1986 | Version v1
Journal article

Field tests of a tissue-equivalent beta survey meter

  • 1. Idaho National Engineering Lab., Idaho Falls (USA)

Description

A tissue-equivalent survey meter for monitoring the (dD/dt)(0.07) and (dD/dt)(10) dose rates produced by betas and photons has been designed and tested. The very thin tissue-equivalent plastic scintillator closely simulates the critical skin tissue layer between 4 mg.cm-2 and 9 mg.cm-2. The meter is calibrated to read the (dD/dt)(0.07) dose rate directly, and the (dD/dt)(10) dose rate behind a filter. Laboratory measurements of calibrated sources and field tests have demonstrated tissue-equivalent response of the survey meter to betas between 70 keV and 2500 keV and to photons with energies greater than about 30 keV. (author)

Additional details

Publishing Information

Journal Title
Radiat. Prot. Dosim.
Journal Volume
14
Journal Issue
2
Series
Radiat. Prot. Dosim.
Journal Page Range
117-121
ISSN
0144-8420
CODEN
RPDOD

Conference

Title
Workshop on dosimetry of beta particles and low energy X-rays.
Dates
7-9 Oct 1985.
Place
Saclay (France).

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
17069775
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BETA DOSIMETRY; FIELD TESTS; GAMMA DOSIMETRY; PERFORMANCE; SURVEY MONITORS; TISSUE-EQUIVALENT DETECTORS
Descriptors DEC
DOSIMETRY; MEASURING INSTRUMENTS; MONITORS; RADIATION DETECTORS; RADIATION MONITORS

Optional Information

Secondary number(s)
CONF--8510185; EUR--10192.