Field tests of a tissue-equivalent beta survey meter
- 1. Idaho National Engineering Lab., Idaho Falls (USA)
Description
A tissue-equivalent survey meter for monitoring the (dD/dt)(0.07) and (dD/dt)(10) dose rates produced by betas and photons has been designed and tested. The very thin tissue-equivalent plastic scintillator closely simulates the critical skin tissue layer between 4 mg.cm-2 and 9 mg.cm-2. The meter is calibrated to read the (dD/dt)(0.07) dose rate directly, and the (dD/dt)(10) dose rate behind a filter. Laboratory measurements of calibrated sources and field tests have demonstrated tissue-equivalent response of the survey meter to betas between 70 keV and 2500 keV and to photons with energies greater than about 30 keV. (author)
Additional details
Publishing Information
- Journal Title
- Radiat. Prot. Dosim.
- Journal Volume
- 14
- Journal Issue
- 2
- Series
- Radiat. Prot. Dosim.
- Journal Page Range
- 117-121
- ISSN
- 0144-8420
- CODEN
- RPDOD
Conference
- Title
- Workshop on dosimetry of beta particles and low energy X-rays.
- Dates
- 7-9 Oct 1985.
- Place
- Saclay (France).
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 17069775
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BETA DOSIMETRY; FIELD TESTS; GAMMA DOSIMETRY; PERFORMANCE; SURVEY MONITORS; TISSUE-EQUIVALENT DETECTORS
- Descriptors DEC
- DOSIMETRY; MEASURING INSTRUMENTS; MONITORS; RADIATION DETECTORS; RADIATION MONITORS
Optional Information
- Secondary number(s)