Task-based assessment on various optimization protocols of computed tomography Pulmonary Angiography examination
Creators
- 1. Department of Physics, Faculty of Science, Universiti Putra Malaysia, 43400, Serdang, Selangor (Malaysia)
- 2. Centre for Applied Physics and Radiation Technologies, School of Engineering and Technology, Sunway University, No 5, Jalan Universiti, Bandar Sunway, 47500, Petaling Jaya, Selangor (Malaysia)
- 3. Department of Physics, College of Natural Sciences, Jeonbuk National University, 567,Baekje-daero, Deokjin-gu, Jeonju-si, Jeollabuk-do, 54896 (Korea, Republic of)
Description
Highlights: • The performance of CTPA examination was assessed using task-based assessment. • The CATPHAN-600 phantom has been utilized and scan under 128-slice CT scanner. • Low dose exposure setting with higher IR level decreases the TTF spatial frequency. • SNR is significantly reduced when a high pitch factor is used. This prospective study was designed to assess CT performance by using the task-based image quality assessment and to identify the influence of iterative reconstruction (IR) algorithm, tube potential and pitch factor selection on dose and image quality during Computed Tomography Pulmonary Angiography (CTPA) examination. An image quality phantom, Catphan-600 was utilized in this study for assessing the image quality of CTPA protocols under 128-multislice CT scanner (Philips Brilliance iCT, USA). The standard CTPA protocols were altered on its tube potential, pitch factor and iterative reconstruction algorithm (IR). Three different levels of iDose, an IR algorithm were measured in this study. The result was presented in term of signal to noise ratio (SNR), contrast to noise ratio (CNR), noise power spectrum (NPS) and target transfer function (TTF). A massive improvement of the CNR and distortion of noise texture were observed with increasing the IR algorithm level. Noise value reduction was significantly achieved by increasing the tube potential selection although there are no changes in TTF values. The alteration of pitch factor provides some fluctuation on both NPS and TTF values. In conclusion, the task-based measurement was performed, and this technique allow for characterizing the image quality with a different type of quantitative measurement for local CTPA protocols.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radphyschem.2021.109692Additional details
Identifiers
- DOI
- 10.1016/j.radphyschem.2021.109692;
- PII
- S0969806X2100342X;
Publishing Information
- Journal Title
- Radiation Physics and Chemistry (1993)
- Journal Volume
- 188
- Journal Page Range
- vp.
- ISSN
- 0969-806X
- CODEN
- RPCHDM
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 54050195
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; BIOMEDICAL RADIOGRAPHY; BLOOD VESSELS; COMPUTERIZED TOMOGRAPHY; FLUCTUATIONS; IMAGES; ITERATIVE METHODS; OPTIMIZATION; PHANTOMS; SIGNAL-TO-NOISE RATIO; TRANSFER FUNCTIONS
- Descriptors DEC
- BODY; CALCULATION METHODS; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; FUNCTIONS; MATHEMATICAL LOGIC; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; STRUCTURAL MODELS; TOMOGRAPHY; VARIATIONS
Optional Information
- Copyright
- Copyright (c) 2021 Elsevier Ltd. All rights reserved.