Measurement of the detective quantum efficiency (DQE) of digital X-ray detectors according to the novel standard IEC 62220-1
Creators
- 1. Physikalisch-Technische Bundesanstalt, Bundesallee 100, D-38116 Braunschweig (Germany)
- 2. Dept. for Diagnostic Radiology, Univ. Hospital Magdeburg, Leipziger Strasse 44, D-39120 Magdeburg (Germany)
Description
A mobile measurement facility which complies with IEC 62220-1 has been set up to determine the detective quantum efficiency (DQE) of digital X-ray detector systems. Exemplary measurements were performed for two similar CR detector systems, a CsI-based indirect detector and an Se-based direct detector. The standardised radiation quality RQA 5 was applied for measurement and for three of these systems RQA 9 was also applied. A pronounced dependence of DQE on radiation quality was observed for the direct detector, where the DQEs for RQA 5 and RQA 9 differ by a factor of ∼2. The uncertainty (95% confidence interval) associated with the measured DQE values is within 0.01 and 0.04 depending on, for example, the spatial frequency. Thus, it has been demonstrated that the DQE can be measured accurately and reliably with the accuracy required by the international standard IEC 62220-1. It is now possible to objectively measure and compare DQE values of digital X-ray detector systems. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/nch507Additional details
Identifiers
- DOI
- 10.1093/rpd/nch507;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 114
- Journal Issue
- 1-3
- Journal Page Range
- p. 39-44
- ISSN
- 0144-8420
Conference
- Title
- 2. Malmoe Conference on Medical X-Ray Imaging - Optimisation strategies in medical X-ray imaging
- Dates
- 23-25 Apr 2004
- Place
- Malmoe (Sweden)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 44035764
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; IMAGES; QUANTUM EFFICIENCY; RADIATION DETECTORS; RADIATION QUALITY; STANDARDIZATION; STANDARDS; X RADIATION
- Descriptors DEC
- EFFICIENCY; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATIONS
Optional Information
- Notes
- 13 refs