Published January 2002 | Version v1
Journal article

Calibration of the NPL secondary standard radionuclide calibrator for 125I seeds used for prostate brachytherapy

Description

In the therapeutic use of radionuclides, by far the most rapid growth in recent years is that of 125I seeds used for the treatment of prostate cancer. Large numbers of these seeds are used in each treatment and there is a need for a simple but accurate means of confirming their dose rates. This mechanism requires a transfer device for which the calibration factors are traceable to national standards. The NPL secondary standard radionuclide calibrator, because of its guaranteed reproducibility and traceable calibration procedure, is ideally suited for this purpose. A series of characterisation measurements have been performed on the NPL radionuclide calibrator in order to estimate the uncertainty levels that can be achieved and these are presented together with the relevant calibration factors for some typical seeds

Additional details

Identifiers

PII
S0969804301002081;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
56
Journal Issue
1-2
Journal Page Range
p. 321-325
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.