Published March 1993 | Version v1
Journal article

An enhanced sensitivity K-shell x-ray fluorescence technique for tibial lead determination

  • 1. Queen Elizabeth Medical Centre, Birmingham (United Kingdom). Dept. of Medical Physics and Biomedical Engineering
  • 2. McMaster Univ., Hamilton, ON (Canada). Dept. of Physics

Description

A novel irradiation-detection geometry capable of enhancing sensitivity for the measurement of tibial lead content by K-shell x-ray fluorescence (XRF) is described. The high-count-rate system comprised a small-area high-specific-activity (0.147 GBq mm-2) 109Cd source and a large-area (nominally 20 cm2) uncollimated detector, forming an axially symmetric back-scattering arrangement. Precisions in the range ±4.9 to ±14.2 μg Pb (g bone mineral)-1 have been obtained in a study of a cohort of 63 controls and 73 workers industrially exposed to lead. These precisions are comparable with those obtained in results using earlier systems, but at reduced source activities (less than 50% of the activity of other systems) and with significant reduction in measurement time (some 30% less than the measurement times of other systems). (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
38
Journal Issue
3
Journal Page Range
p. 389-396.
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
24048920
Subject category
S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
Descriptors DEI
BONE TISSUES; IN VIVO; K SHELL; LEAD; TIBIA; X-RAY FLUORESCENCE ANALYSIS
Descriptors DEC
BODY; CHEMICAL ANALYSIS; CONNECTIVE TISSUE; ELECTRONIC STRUCTURE; ELEMENTS; METALS; NONDESTRUCTIVE ANALYSIS; ORGANS; SKELETON; TISSUES; X-RAY EMISSION ANALYSIS