Published July 1986 | Version v1
Journal article

A dish for determining tritium content with liquid radiometer

Description

The results of experimental studies on selecting material and optimum geometrical dimensions of a measuring dish of a high-sensitivity liquid radiometer designed for mass analysis of tritium in the form of tritiated water in environmental aquatic objects at the 104 Bq/m3 level are presented. The experiments are carried out in the detection unit using the FEhU-118 photomultiplier with semitransparent antimony-potassium-sodium-cesium photocathodes characterized by light sensitivity of (1-2)x10-4 A/lm and measurement stability. It is shown that a dish made of polytetrafluorethylene with wall thickness of 0.8-1.0 mm and useful capacity of 45 cm3 (5 cm3 of tritiated water and 40 cm3 of liquid scintillator ZhS-8I) is chosen as an optimum one

Additional details

Additional titles

Original title (Russian)
Кювета для определения количества трития жидкостным радиометром

Publishing Information

Journal Title
Prib. Tekh. Ehksp.
Journal Issue
no.4
Series
Prib. Tekh. Ehksp.
ISSN
0032-8162
CODEN
PRTEA

Optional Information

Notes
For English translation see the journal Instruments and Experimental Techniques (USA).