CT angiography using electron-beam computed tomography (EBCT). A phantom study
Description
The purpose of this study was to evaluate the accuracy of CT angiography in small vessels using electron-beam computed tomography (EBCT). Vessel phantoms with inner diameters of 8 mm, 6 mm, and 4 mm were prepared with segments of 75%, 50%, and 25% stenosis in each vessel. The vessels were filled with contrast medium (Iopamidol 300 at 1/24 dilution, approximately 380 HU). The EBCT apparatus used was an Imatron C-150. The step volume scan mode was used with slice thicknesses of 1.5 mm and 3.0 mm, scan time of 0.3 sec, and 210 mm field of view. Images with a slice thickness of 1.5 mm were definitely better than those with a slice thickness of 3.0 mm. The quality of maximum intensity projection (MIP) images was quite similar to that of three-dimensional (3D) images. Using the 8 mm vessel phantom, all stenotic segments were accurately visualized on CT angiography. The 50% stenotic segments were accurately estimated in all vessels. However, the 75% stenotic segments were slightly overestimated in smaller vessels, and the 25% stenotic segments were slightly underestimated in smaller vessels. We consider CT angiography using EBCT to be a useful, less invasive diagnostic modality for stenoocclusive lesions. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Medicine (Tokyo)
- Journal Volume
- 15
- Journal Issue
- 5
- Journal Page Range
- p. 273-276.
- ISSN
- 0288-2043
- CODEN
- RAMEER
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 29019282
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; BLOOD VESSELS; CAT SCANNING; IMAGES; IOPAMIDOL; PHANTOMS; THREE-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- BODY; CARDIOVASCULAR SYSTEM; COMPUTERIZED TOMOGRAPHY; CONTRAST MEDIA; MOCKUP; ORGANS; STRUCTURAL MODELS; TOMOGRAPHY