Multicomponent measurements of radiation doses received by patients undergoing ct-examination with three type TLDs

  • 1. Inter-University Centre on Medical Radiation Physics, Plovdiv (Bulgaria)
  • 2. Medical University, Department of Roentgeology, Plovdiv (Bulgaria)
  • 3. Medical University, Department of Medical Physicsand Biophysics, Plovdiv (Bulgaria)
  • 4. Plovdiv University, Department of Atomic Physics, Plovdiv (Bulgaria)

Description

About 20 % of the radiation dose during the radiological examinations is received from CT. The determination of the entrance surface dose is an important part of QA. Three types of TLD: LiF, CaSO4:Dy and α-Al2O3:C have been used after selection for homogeneity and reproducibility (SD<10 %) and calibration on the phantom surface. A large variation in the sensitivity in TLD responses from different luminescence materials is demonstrated. A good correlation coefficient (0,98) between CTDT and TLD-responses was observed. A space dose distribution profiles as well as the surface doses were determined for the center of the single scan and periphery in the gonad and thyroid gland positions (MU-Plovdiv). (author)

Part of:
Scientific Proceedings of Riga Technical University

Additional details

Additional titles

Original title (English)
Rigas Tehniskas Universitates Zinatniskie Raksti

Publishing Information

Publisher
RTU
Imprint Place
Riga (Latvia)
Imprint Title
Scientific Proceedings of Riga Technical University
Imprint Pagination
198 p.
Journal Volume
9. Biomedical Engineering and Microtechnologies
Series
Series 6. Mashine Science and Transport
Journal Page Range
p. 62-65
ISSN
1407-8015

Optional Information

Notes
8 refs., tab., fig., e-mail: nasko4444@abv.bg Imprint:Rigas Tehniskas Universitates Zinatniskie Raksti