The UV absorption of graphene oxide is size-dependent: possible calibration pitfalls
Creators
- 1. Sichuan University, Department of Pharmacy, West China Hospital (China)
- 2. Sichuan University, State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology (China)
Description
Graphene oxide (GO) is often quantified via its UV absorption, typically at around 230 nm. This is convenient but the effect of the size of GO on the accuracy of this method has been ignored so far. The authors report that the molar absorbance of GO is size-dependent. Data are presented on the absorbance of small (hydrodynamic diameter 1 μm), medium sized (1.5 μm), and large (2.2 μm) GO particles at wavelengths of 210, 230 and 250 nm. In general, linear relationship and good regression fits are obtained, but with different slope depending on size even at the same wavelength. This implies that using the UV absorption-based calibration may cause significant errors in GO quantification. Ultimately, this leads to incorrect dosages and faulty conclusions. This may also explain a variety of inconsistent results obtained in previous biological applications of GO.
.
Additional details
Identifiers
Publishing Information
- Journal Title
- Mikrochimica Acta
- Journal Volume
- 186
- Journal Issue
- 3
- Journal Page Range
- p. 1-7
- ISSN
- 0026-3672
- CODEN
- MIACAQ
INIS
- Country of Publication
- Austria
- Country of Input or Organization
- Austria
- INIS RN
- 51103134
- Subject category
- S77: NANOSCIENCE AND NANOTECHNOLOGY; S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- ACCURACY; BIOASSAY; CALIBRATION; ERRORS; GRAPHENE; NANOCHEMISTRY; NANOSTRUCTURES; SIZE; ULTRAVIOLET RADIATION; ULTRAVIOLET SPECTRA; WAVELENGTHS
- Descriptors DEC
- CARBON; CHEMISTRY; ELECTROMAGNETIC RADIATION; ELEMENTS; NONMETALS; RADIATIONS; SPECTRA
Optional Information
- Copyright
- Copyright (c) 2019 Springer-Verlag GmbH Austria, part of Springer Nature