Published December 2000
| Version v1
Journal article
Offline calibration of BES II vertex chamber
Creators
- 1. Academia Sinica, Beijing (China). Inst. of High Energy Physics
Description
An algorithm based on VCJULI to calibrate the BES II vertex chamber is described. The calibration includes the determination of global T0, the subtraction of t0 shift for each wire, the corrections of wire position itself and the position of the vertex chamber relative to the main drift chamber, the establishment of drift time to drift distance relation, etc. The main performance of the BES II vertex chamber is given accordingly: space resolution is better than 100 μm
Additional details
Publishing Information
- Journal Title
- High Energy Physics and Nuclear Physics
- Journal Volume
- 24
- Journal Issue
- 12
- Journal Page Range
- p. 1165-1172
- ISSN
- 0254-3052
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 32018557
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ALGORITHMS; BEIJING ELECTRON-POSITRON COLLIDER; CALIBRATION; CORRECTIONS; MAGNETIC SPECTROMETERS; RADIATION DETECTORS; SPATIAL RESOLUTION
- Descriptors DEC
- ACCELERATORS; LINEAR ACCELERATORS; MEASURING INSTRUMENTS; RESOLUTION; SPECTROMETERS; STORAGE RINGS