Detection of thoracic vascular structures by electrical impedance tomography: a systematic assessment of prominence peak analysis of impedance changes

  • 1. Department of Anesthesiology, Center of Anaesthesiology and Intensive Care Medicine, University Medical Center Hamburg-Eppendorf, Hamburg (Germany)
  • 2. Swisstom AG, Schulstrasse 1, 7302 Landquart (Switzerland)
  • 3. Texcess, Hamburg (Germany)
  • 4. Institute of Electrodynamics, Microwave and Circuit Engineering, Vienna University of Technology, Gußhausstraße 25/354, 1040 Vienna (Austria)
  • 5. Department of Anesthesiology and Intensive Care Medicine, University Medical Center Rostock, Rostock (Germany)

Description

Objective: Electrical impedance tomography (EIT) is a non-invasive and radiation-free bedside monitoring technology, primarily used to monitor lung function. First experimental data shows that the descending aorta can be detected at different thoracic heights and might allow the assessment of central hemodynamics, i.e. stroke volume and pulse transit time. Approach: First, the feasibility of localizing small non-conductive objects within a saline phantom model was evaluated. Second, this result was utilized for the detection of the aorta by EIT in ten anesthetized pigs with comparison to thoracic computer tomography (CT). Two EIT belts were placed at different thoracic positions and a bolus of hypertonic saline (10 ml, 20%) was administered into the ascending aorta while EIT data were recorded. EIT images were reconstructed using the GREIT model, based on the individual's thoracic contours. The resulting EIT images were analyzed pixel by pixel to identify the aortic pixel, in which the bolus caused the highest transient impedance peak in time. Main results: In the phantom, small objects could be located at each position with a maximal deviation of 0.71 cm. In vivo, no significant differences between the aorta position measured by EIT and the anatomical aorta location were obtained for both measurement planes if the search was restricted to the dorsal thoracic region of interest (ROIs). Significance: It is possible to detect the descending aorta at different thoracic levels by EIT using an intra-aortic bolus of hypertonic saline. No significant differences in the position of the descending aorta on EIT images compared to CT images were obtained for both EIT belts. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6579/aaa924

Additional details

Identifiers

Publishing Information

Journal Title
Physiological Measurement (Print)
Journal Volume
39
Journal Issue
2
Journal Page Range
[8 p.]
ISSN
0967-3334

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
51083994
Subject category
S60: APPLIED LIFE SCIENCES;
Resource subtype / Literary indicator
Numerical Data
Descriptors DEI
AORTA; COMPUTERIZED TOMOGRAPHY; EXPERIMENTAL DATA; IMPEDANCE; IN VIVO; LUNGS; PHANTOMS; SWINE
Descriptors DEC
ANIMALS; ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DATA; DIAGNOSTIC TECHNIQUES; DOMESTIC ANIMALS; INFORMATION; MAMMALS; MOCKUP; NUMERICAL DATA; ORGANS; RESPIRATORY SYSTEM; STRUCTURAL MODELS; TOMOGRAPHY; VERTEBRATES