estimation of Hp(3) to the eye lens of interventional radiologists-relation between the eye lens dose and radiologist's height
- 1. Department of Quantum Medical Technology, Division of Health Sciences, Graduate School of Medical Science, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 920-0942 (Japan)
- 2. Department of Clinical Radiology, Tottori University Hospital, 36-1 Nishi-cho, Yonago, Tottori 683-8504 (Japan)
- 3. Division of Diagnostic Medical Physics, Department of Radiology, Virginia Commonwealth University Medical Center, 1300 East Marshall Street, Richmond, VI 23298-0615 (United States)
- 4. Preparing Section for New Faculty of Medical Science, Fukushima Medical University, 1 Hikariga-oka, Fukushima, Fukushima 960-1295 (Japan)
Description
The aim of the study was to estimate occupational radiation dose to the eye lens of radiologists and the dose reduction ratio of lead glasses during interventional radiology. Three interventional radiologists monitored Hp(3) using small-type optically stimulated luminescence dosemeters attached to the left inside and outside of the lead glasses with 0.07-mmPb [Hp(3)eye]. Hp(10) and Hp(0.07) were monitored, respectively, by attaching the personal dosemeter to the lead neck collar above the lead apron. The median Hp(3)eye with lead glasses and the median dose reduction ratio of lead glasses for the three radiologists were 8.02 mSv/y and 57.7%, respectively. The median Hp(3)eye without lead glasses [Hp(3)eye-w/o] for the three radiologists was 18.6 mSv/y, but Hp(3)eye-w/o for one of the radiologists was 24.1 mSv/y. Monitoring occupational radiation dose to the eye lens is important because interventional radiologists are at risk of exceeding the new dose limit. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncz181Additional details
Identifiers
- DOI
- 10.1093/rpd/ncz181;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 187
- Journal Issue
- 4
- Journal Page Range
- p. 409-417
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 51108201
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; CRYSTALLINE LENS; DOSE LIMITS; DOSEMETERS; GLASS; HEALTH HAZARDS; HEIGHT; LUMINESCENCE; NECK; RADIATION DOSES; RADIATION MONITORING
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DIMENSIONS; DOSES; EMISSION; EYES; FACE; HAZARDS; HEAD; MEASURING INSTRUMENTS; MEDICINE; MONITORING; NUCLEAR MEDICINE; ORGANS; PHOTON EMISSION; RADIOLOGY; SAFETY STANDARDS; SENSE ORGANS; STANDARDS
Optional Information
- Notes
- 22 refs.