Published November 15, 1993 | Version v1
Journal article

Nonlinearity at the K-absorption-edge in the Xe-filled gas proportional counter

  • 1. Dept. of Earth and Space Science, Faculty of Science, Osaka Univ. (Japan)
  • 2. Dept. of Physics, Faculty of Science, Kyoto Univ. (Japan)

Description

We measured the linearity of a Xe-filled gas proportional counter in the energy range between 10 and 40 keV at the synchrotron radiation facility. A clear gap at the XeK-edge energy (34.5 keV) can be seen. We found that the equivalent width of this gap is about 170±10 eV. In order to understand this value, we discussed the mean ionization energy during the X-ray photo absorption process. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
336
Journal Issue
1-2
Journal Page Range
p. 301-303.
ISSN
0168-9002
CODEN
NIMAER