Published June 2021 | Version v1
Journal article

Surface dose estimation by a KAP meter for kilovoltage X-ray beams

  • 1. Department of Health Sciences, Faculty of Life Sciences, Kumamoto University, 4-24-1 Kuhonji, Chuo-ku, Kumamoto 862-0976 (Japan)
  • 2. SAGA Heavy Ion Medical Accelerator in Tosu, 3049 Harakogamachi, Tosu, Saga, 841-0071 (Japan)

Description

This study aims to estimate the entrance surface dose (ESD) of a water phantom for kilovoltage x-ray beams using an air kerma area product meter (KAP meter) equipped in an x-ray unit. The KAP meter was calibrated in terms of the ESD determined by a plane-parallel ionization chamber based on a 60Co absorbed dose-to-water calibration coefficient, N60CoD,w. The ESD measured using the KAP meter was verified by comparing it with that estimated by the air kerma calibration coefficient, NK, for x-ray beam qualities. The ratio of ESDs based on N60CoD,w and NK was 1.003 on average and independent of the beam quality. The ESD by the KAP meter was an agreement within ±1.5% with that measured using the plane-parallel chamber for 10 x 10-30 x 30 cm2 fields with a source-surface distance of 75-150 cm. It was possible to estimate the ESD directly in a water phantom for x-ray beams without correction factors compared to the existing air kerma calibration, using a KAP meter calibrated based on N60CoD,w. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncab122

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
195
Journal Issue
2
Journal Page Range
p. 124-131
ISSN
0144-8420

Optional Information

Notes
19 refs.