Temperature cycling test of planar hyper-pure germanium radiation detector
Creators
- 1. Japan Atomic Energy Research Inst., Tokai, Ibaraki. Tokai Research Establishment
Description
If a Ge (Li) detector is left at the normal temperature, generally it does not recover its original performance even when it is cooled again with liquid nitrogen, as Li ions in the compensated i zone precipitate by Li drift and it returns to p type which is the state before drift. One of the devices that overcomes this shortcoming is the p-n junction Ge detector, which required the production of high purity Ge single crystals to obtain the thick depletion layer. The planar or coaxial type detectors were produced using the Ge single crystals with impurity concentration of 1010/cm3 and it was recognized that they showed the gamma detecting characteristic nearly equal to Ge (Li) detectors. They are now commercially available from a few companies. The author carried out the temperature-cycling test of the planar type hyperpure Ge detector sold by Nuclear Radiation Developments, Canada. First, applying liquid nitrogen, the leakage current, static capacity, gamma ray-detecting efficiency and energy resolution were measured. Then it was returned to room temperature. Since then, irregular cycling tests were carried out 15 times. The results didn't show any significant change in the gamma ray-detecting efficiency, energy resolution and static capacity. Though leakage current changed between 9.3 and 33 pA, it does not influence on the energy resolution because of small absolute values. It may be said that it is sufficiently stable in the temperature cycling from room temperature to 77 K. (Wakatsuki, Y.)
Files
Additional details
Publishing Information
- Imprint Title
- Study-meeting report on improvements of accuracy and speed in #gamma#-ray measurements
- Journal Page Range
- p. 10-12.
- Report number
- KURRI-TR--142
Conference
- Title
- Meeting on improvements of accuracy and speed in #gamma#-ray measurements.
- Dates
- 19 Jan 1976.
- Place
- Kumatori, Osaka, Japan.
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 8344116
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ELECTROSTATICS; ENERGY RESOLUTION; GAMMA SPECTROSCOPY; HIGH-PURITY GE DETECTORS; LEAKAGE CURRENT; PERFORMANCE TESTING; THERMAL CYCLING
- Descriptors DEC
- CURRENTS; ELECTRIC CURRENTS; GE SEMICONDUCTOR DETECTORS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RESOLUTION; SEMICONDUCTOR DETECTORS; SPECTROSCOPY; TESTING