Published July 1984
| Version v1
Journal article
High-temperature specimen holder for use with the PHI model 2100 specimen introduction/reaction chamber
Creators
- 1. Metals and Ceramics Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831
Description
A high-temperature specimen holder for use with a commercially available rapid specimen introduction device on an Auger electron spectroscopy system is described. Necessary (nondestructive) modifications to the specimen carousel are also included. Use of the specimen holder allows rapid transfer of specimens for in situ heating to approximately 10000 C during Auger analysis, while retaining conventional capabilities of the commercial system for ambient temperature analysis
Additional details
Publishing Information
- Journal Title
- Rev. Sci. Instrum.
- Journal Volume
- 55
- Journal Issue
- 7
- Series
- Rev. Sci. Instrum.
- Journal Page Range
- 1171-1172
- ISSN
- 0034-6748
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 16011217
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- AUGER ELECTRON SPECTROSCOPY; CATALYSIS; IN-SITU PROCESSING; SAMPLE HOLDERS; SEGREGATION; SURFACES; USES; VERY HIGH TEMPERATURE
- Descriptors DEC
- ELECTRON SPECTROSCOPY; ORE PROCESSING; SPECTROSCOPY