Published October 2006 | Version v1
Journal article

Visible spectrometer at the Compact Toroid Injection Experiment and the Alcator C-Mod tokamak for Doppler width and shift measurements

  • 1. Plasma Science and Fusion Center, Massachusetts Institute of Technology, 190 Albany Street, Cambridge, Massachusetts, 02139 (United States)
  • 2. Lawrence Livermore National Laboratory, 7000 East Avenue, Livermore, California 94550 (United States)
  • 3. University of California at Davis, One Shields Avenue, Davis, California 95616 (United States)

Description

A novel Doppler spectrometer is currently being used for ion or neutral velocity and temperature measurements on the Alcator C-Mod tokamak. The spectrometer has an f∼3.1 and is appropriate for visible light (3500-6700 A). The linewidth from a line emitting calibration source has been measured to be as small as 0.4 A ring . The ultimate time resolution is line brightness light limited and on the order of milliseconds. A new photon efficient charge coupled device detector is being used at C-Mod. Time resolution is achieved by moving the camera during a plasma discharge in a perpendicular direction through the dispersion plane of the spectrometer, causing a vertical streaking across the camera face. Initial results from C-Mod as well as previous measurements from the Compact Toroid Injection Experiment are presented

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
77
Journal Issue
10
Journal Page Range
p. 10F125
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2006 American Institute of Physics