Published November 1977 | Version v1
Journal article

Radiation from the uranium collimators of the Varian 4-MeV accelerator

  • 1. Department of Radiotherapy, Howard University Hospital, College of Medicine, Washington, DC 20060

Description

The uranium collimator of the Varian 4-MeV accelerator produces a dose rate of 100 mR/h near the collimator opening. It can be decreased to 4% by a 6-mm Lucite shield at the level of the collimator opening. A better solution is a leadglass--mylar (LM) shield which decreases the uranium radiation to less than 3% and also provides better skin sparing during treatment than the open collimator

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
4
Journal Issue
6
Series
Med. Phys.
Journal Page Range
526-527
ISSN
0094-2405

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
9374074
Subject category
S43: PARTICLE ACCELERATORS;
Descriptors DEI
COLLIMATORS; DOSE RATES; LINEAR ACCELERATORS; MEV RANGE 01-10; SHIELDING; URANIUM
Descriptors DEC
ACCELERATORS; ACTINIDES; CYCLIC ACCELERATORS; ELEMENTS; ENERGY RANGE; METALS; MEV RANGE

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent