Published March 1988 | Version v1
Journal article

Diagnostic system for passive charge-exchange particle measurements on JT-60

  • 1. Japan Atomic Energy Research Inst., Naka, Ibaraki. Naka Fusion Research Establishment

Description

In order to measure energy distributions of the charge-exchange neutral particles in the JT-60 experiments, a compact size electrostatic energy analyzer which the measurable energy range is from 1 keV to 100 keV is developed successfully. Compactness of an analyzer is accomplished by setting an accelerator between a gas stripping cell and a deflector of 45deg injection type. The calibration of the analyzer was carried out owing to confirm the capability of energy analysis and stripping efficiency. This analyzer was applied to measure the energy distribution in additionally heated plasmas in JT-60. The usefullness of the analyzer was confirmed. (author)

Additional details

Publishing Information

Journal Title
Kaku Yugo Kenkyu
Journal Volume
59
Journal Issue
suppl
Series
Kaku Yugo Kenkyu.
Journal Page Range
91-118
ISSN
0451-2375
CODEN
KAKEA