Published July 21, 2009 | Version v1
Journal article

Absorption measurements on a new cone beam CT and IMRT compatible tabletop for use in external radiotherapy

  • 1. Department of Medical Physics, Vejle Hospital, DK-7100 Vejle (Denmark)
  • 2. Department of Radiation Oncology, Rigshospitalet, Copenhagen University Hospital, Copenhagen (Denmark)

Description

With the development of intensity-modulated radiotherapy (IMRT), intensity-modulated arc therapy (IMAT), volumetric single arc therapy (RapidArc) and cone beam CT (CBCT) in 3D conformal radiotherapy, a need for new homogeneous treatment tabletops has evolved. Therefore, a new homogenous carbon fibre rotatable treatment tabletop Contesse has been developed. The intension was to create a CT compatible treatment tabletop accommodating all the standard treatment areas without add-on devises or the need to reconfigure the tabletop in everyday practice. At the same time, the absorption was sought to be minimized. Measurements of absorption with an ionization chamber have been performed for the main treatment geometries at normal and oblique angles together with build-up measurements at normal incidence. Absorption and build-up properties were shown to be comparable to those of other treatment tabletops with similar fixation devices. No artefacts from the tabletop are seen on CBCT scans. In order to reveal the absorption characteristics of tabletops, careful measurements at oblique angles including fixation devices on all clinically used tabletops must be performed. The Contesse tabletop is CBCT compatible, has acceptable absorption and is presently being used in the clinic for all patients, including IMRT and kilovoltage CBCT. (note)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/54/14/N04

Additional details

Identifiers

DOI
10.1088/0031-9155/54/14/N04;
PII
S0031-9155(09)08026-9;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
54
Journal Issue
14
Journal Page Range
p. N319-N328
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41050866
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ABSORPTION; CARBON FIBERS; COMPUTERIZED TOMOGRAPHY; IONIZATION CHAMBERS; RADIOTHERAPY
Descriptors DEC
DIAGNOSTIC TECHNIQUES; FIBERS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIOLOGY; SORPTION; THERAPY; TOMOGRAPHY