Measurement of Long-Range Interatomic Distances by Solid-State Tritium-NMR Spectroscopy

  • 1. CEA, iBiTecS, SerVice de Chimie Bioorganique et de Marquage, F-91191 Gif sur YVette, (France)
  • 2. CEA, iBiTecS, SerVice de Bioenergetique, Biologie Structurale et Mecanismes, F-91191 Gif sur YVette, (France)
  • 3. CEA, IRAMIS, SIS2M, Laboratoire Structure et Dynamique par Resonance Magnetique, F-91191 Gif sur YVette, (France)
  • 4. LBPA, CNRS, ENS de Cachan, 61, aVenue du President Wilson, F-94235 Cachan, (France)

Description

In summary, we have described a simple, sensitive, and accurate approach for the determination of short- to long-range interatomic distances using standard probe electronics and sample holders and have successfully applied it to model compounds. The measured distance of 14.4±2.2 Angstroms reported here is the highest ever obtained using NMR. To generalize this approach to samples with a larger 3H chemical shift distribution or shorter T1ρ values, modifications of the experimental conditions such as use of higher spinning frequency, sample volume reduction, or design of new pulse sequences would be beneficial. (authors)