Published February 1, 1992 | Version v1
Journal article

The HPGe as a defined-solid-angle detector for low-energy photons

  • 1. Australian Radiation Lab., Yallambie (Australia)

Description

HPGe detectors now available are well suited for use as defined-solid-angle spectrometers in the low-energy range, 25 to 80 keV. The calculation of efficiency, as defined by an external collimator, to very high accuracy seems possible if account is taken of the various mechanisms through which photons can be lost from the full-energy peak. Compton scattering is important in this energy range and its effect on peak shape and detection efficiency has been investigated for the 59.5 keV γ-ray from 241Am and compared with calculation. The close agreement found suggests that an uncertainty of less than 1% is attainable. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A
Journal Volume
312
Journal Issue
1/2
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
146-151
ISSN
0168-9002
CODEN
NIMAE

Conference

Title
Biennial meeting of the International Committee for Radionuclide Metrology (ICRM) including the international symposium on radionuclide metrology and its applications.
Dates
27-31 May 1991.
Place
Madrid (Spain).