Measurement of gamma ray dose from a moderated 252Cf source
- 1. Battelle Pacific Northwest Labs., Richland, WA (USA)
- 2. Lawrence Livermore National Lab., CA (USA)
- 3. Michigan Univ., Ann Arbor (USA)
Description
The fraction of dose equivalent due to gamma rays from a moderated 252Cf source was determined using three types of dosimetry system. Two A-150 tissue-equivalent (TE) plastic ionisation chambers and a TE proportional counter (TEPC) were used to measure tissue dose, from which the neutron dose equivalent was computed. The ratio of gamma ray dose equivalent to the neutron dose equivalent was determined by using a relatively neutron insensitive Geiger-Mueller (GM) counter and thermoluminescence dosemeters (TLD). In addition, the event-size spectrum measured by the TEPC was also used to compute the gamma ray dose fraction. The average value of the ratio of gamma ray dose to neutron dose equivalent was found to be 0.18 with an uncertainty of about +- 18%. (author)
Additional details
Publishing Information
- Journal Title
- Radiat. Prot. Dosim.
- Journal Volume
- 9
- Journal Issue
- 2
- Series
- Radiat. Prot. Dosim.
- Journal Page Range
- 113-118
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 16030694
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- CALIFORNIUM 252; COMPARATIVE EVALUATIONS; DOSE EQUIVALENTS; GAMMA DOSIMETRY; NEUTRON DOSIMETRY; NEUTRON SOURCES
- Descriptors DEC
- ACTINIDE NUCLEI; ALPHA DECAY RADIOISOTOPES; CALIFORNIUM ISOTOPES; DOSIMETRY; EVEN-EVEN NUCLEI; HEAVY NUCLEI; ISOTOPES; NUCLEI; PARTICLE SOURCES; RADIATION SOURCES; RADIOISOTOPES; YEARS LIVING RADIOISOTOPES