Published 1984
| Version v1
Report
Open
Low pressure mucltiste avalanche chamber performance
Description
A multistep low pressure avalanche chamber is described. The chamber operates using isobutane at a pressure of 6 torr. The area of the chamber is 240x180 mm. The detector has a high gas amplification coefficient of 3x106 at a nonuniformity of +-3% over the area. The space resolution for both coordinates is 170 μm (FWHM) for isotropic α-radiators placed in the immediate vicinity of a preamplifier gap of 0.5 mm. The good uniformity of the amplification coefficient, low noises and high stability make it possible to use such a coordinate detector in nuclear physics and for solving applied problems, e. g., in neutron radiography
Availability note (English)
MF available from INIS under the Report Number.Files
16054842.pdf
Files
(247.4 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:7e9129fadbbe48d3e7b5b24235279aa3
|
247.4 kB | Preview Download |
System files
(21.0 kB)
| Name | Size | Download all |
|---|
Additional details
Additional titles
- Original title (Russian)
- Характеристики многоступенчатой лавинной камеры низкого давления
Publishing Information
- Imprint Pagination
- 6 p.
- Report number
- JINR-R--13-84-563
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 16054842
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ELECTRIC FIELDS; LOW PRESSURE; MULTIWIRE PROPORTIONAL CHAMBER; NEUTRON RADIOGRAPHY; PERFORMANCE; POSITION SENSITIVE DETECTORS; SPATIAL RESOLUTION; TOWNSEND DISCHARGE; 2-METHYLPROPANOL
- Descriptors DEC
- ALCOHOLS; ELECTRIC DISCHARGES; HYDROXY COMPOUNDS; MEASURING INSTRUMENTS; ORGANIC COMPOUNDS; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RESOLUTION
Optional Information
- Notes
- 7 refs.; 6 figs.; submitted to the journal Instrum. Exp. Tech.