Published August 21, 2010 | Version v1
Journal article

Verification measurements and clinical evaluation of the iPlan RT Monte Carlo dose algorithm for 6 MV photon energy

  • 1. Radiotherapy Centre West, PO Box 432, NL-2501 CK, The Hague (Netherlands)

Description

This study presents data for verification of the iPlan RT Monte Carlo (MC) dose algorithm (BrainLAB, Feldkirchen, Germany). MC calculations were compared with pencil beam (PB) calculations and verification measurements in phantoms with lung-equivalent material, air cavities or bone-equivalent material to mimic head and neck and thorax and in an Alderson anthropomorphic phantom. Dosimetric accuracy of MC for the micro-multileaf collimator (MLC) simulation was tested in a homogeneous phantom. All measurements were performed using an ionization chamber and Kodak EDR2 films with Novalis 6 MV photon beams. Dose distributions measured with film and calculated with MC in the homogeneous phantom are in excellent agreement for oval, C and squiggle-shaped fields and for a clinical IMRT plan. For a field with completely closed MLC, MC is much closer to the experimental result than the PB calculations. For fields larger than the dimensions of the inhomogeneities the MC calculations show excellent agreement (within 3%/1 mm) with the experimental data. MC calculations in the anthropomorphic phantom show good agreement with measurements for conformal beam plans and reasonable agreement for dynamic conformal arc and IMRT plans. For 6 head and neck and 15 lung patients a comparison of the MC plan with the PB plan was performed. Our results demonstrate that MC is able to accurately predict the dose in the presence of inhomogeneities typical for head and neck and thorax regions with reasonable calculation times (5-20 min). Lateral electron transport was well reproduced in MC calculations. We are planning to implement MC calculations for head and neck and lung cancer patients.

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/55/16/S13

Additional details

Identifiers

DOI
10.1088/0031-9155/55/16/S13;
PII
S0031-9155(10)41639-5;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
55
Journal Issue
16
Journal Page Range
p. 4601-4614
ISSN
0031-9155
CODEN
PHMBA7

Conference

Title
2. european workshop on Monte Carlo treatment planning
Acronym
MCTP2009
Dates
19-21 Oct 2009
Place
Cardiff, Wales (United Kingdom)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
42025258
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ALGORITHMS; CHEST; HEAD; IONIZATION CHAMBERS; LUNGS; MONTE CARLO METHOD; NECK; NEOPLASMS; PHOTONS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; SKELETON; VERIFICATION
Descriptors DEC
BODY; BOSONS; CALCULATION METHODS; DISEASES; DOSES; ELEMENTARY PARTICLES; MASSLESS PARTICLES; MATHEMATICAL LOGIC; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIATION DETECTORS; RADIOLOGY; RESPIRATORY SYSTEM; THERAPY