Published October 2012 | Version v1
Journal article

High resolution ion Doppler spectroscopy at Prairie View Rotamak

  • 1. Solar Observatory Department, Prairie View A and M University, Prairie View, Texas 77446 (United States)
  • 2. Department of Physics, West Virginia University, Morgantown, West Virginia 26506 (United States)

Description

A fast ion Doppler spectroscopy (IDS) diagnostic system is installed on the Prairie View Rotamak to measure ion temperature and plasma flow. The diagnostic employs a single channel photomultiplier tube and a Jarrell-Ash 50 monochromator with a diffraction grating line density of 1180 lines/mm, which allows for first order spectra of 200–600 nm. The motorized gear of the monochromator allows spectral resolution of 0.01 nm. Equal IDS measurements are observed for various impurity emission lines of which carbon lines exhibit stronger intensities. Furthermore, the diagnostics is examined in an experiment where plasma experiences sudden disruption and quick recovery. In this case, the IDS measurements show ∼130% increase in ion temperature. Flow measurements are shown to be consistent with plasma rotation.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
83
Journal Issue
10
Journal Page Range
p. 10D506-10D506.3
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2012 American Institute of Physics