Description
The aim of this study was to evaluate the suitability of size specific dose estimates (SSDE) to estimate patient dose in Fast kVp switching dual energy CT. An anthropomorphic phantom (RAN-110) was repeatedly scanned (chest, abdomen and the pelvis) using a 64 detector row MDCT (Discovery CT750 HD, GE Healthcare, Milwaukee, WI, USA) with various CT parameters, including Fast kVp switching. Dosimetry was performed using thermo-luminescent dosimeters, positioned both superficially and within the phantom. SSDE was calculated for all slices of the anthropomorphic phantom using both the localiser and axial images. In Fast kVp switching, SSDE underestimated the measured absorbed dose for the chest/abdomen region ∼35% at the maximum, but were in closer agreement for the pelvic region about within 10%. In single energy techniques, SSDE could not be applied in the estimation of organ doses, but in Fast kVp switching dual energy techniques, SSDE could be applied for anatomical regions with larger thicknesses. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncy022Additional details
Identifiers
- DOI
- 10.1093/rpd/ncy022;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 181
- Journal Issue
- 3
- Journal Page Range
- p. 261-268
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 49106125
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABDOMEN; ABSORBED RADIATION DOSES; CHEST; COMPUTERIZED TOMOGRAPHY; DOSEMETERS; DOSIMETRY; EVALUATION; IMAGES; LUMINESCENCE; ORGANS; PATIENTS; PELVIS; PHANTOMS; THICKNESS
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DIMENSIONS; DOSES; EMISSION; MEASURING INSTRUMENTS; MOCKUP; PHOTON EMISSION; RADIATION DOSES; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- 12 refs.