Published 1977 | Version v1
Book

Properties of thermoluminescence dosemeters after high gamma irradiation

  • 1. Karlsruhe Nuclear Research Center, Germany
  • 2. Bhabha Atomic Research Centre, Bombay (India). Health Physics Div.

Description

With respect to an application of TLD systems in the high dose range, the dosimetric properties of some commercial TL materials, i.e. LiF, Li2B4O7, CaF2 and CaSO4 with different activators, have been investigated after gamma pre-exposures up to 2.106 R. It is discussed the non-linearity of response, the residual dosemeter reading and the change in response after high pre-exposure

Additional details

Publishing Information

Publisher
A. Scharmann.
Imprint Place
Sao Paulo
Imprint Title
Proceedings of the Fifth International Conference on Luminescence Dosimetry
Imprint Pagination
497 p.
Journal Page Range
p. 94-102.

Conference

Title
5. International Conference on Luminescence Dosimetry.
Dates
14 - 17 Feb 1977.
Place
Sao Paulo, Brazil.