Published 2009 | Version v1
Miscellaneous

Evaluation and comparison of digital methods of photopeak integration in gamma-ray spectrometry

  • 1. Nuclear Science and Technology Research Institute (NSTRI), Atomic Energy Organisation of Iran (Iran, Islamic Republic of)
  • 2. Faculty of Engineering and Technology, Science and Research Campus, Islamic Azad University (Iran, Islamic Republic of)

Description

Full text: A study of the various methods of gamma-ray photopeak computation has been carried out. The Total Peak Area (TPA) method, the methods proposed by Wasson, Covell, Sterlinski, Quittner and Czauderna and some modifications of these methods have been considered. Totally, seven methods have been considered in theoretical study and then practical evaluation has been carried out for six of them (except of Quittner method). As a result of practical evaluation, such as the other studies, the Wasson-Sterlinski method which is the combination of two methods is proposed as the proposed method. However, a modification by Wasson of the total peak area method is considered to be the most advantageous because of its simplicity and the relatively high precision obtained. (author)

Availability note (English)

Available in abstract form only, full text entered in this record

Additional details

Publishing Information

Imprint Pagination
1 p.

Conference

Title
International Conference on Neutron and X-ray Scattering 2009
Acronym
ICNX 2009
Dates
29 Jun - 1 Jul 2009
Place
Kuala Lumpur (Malaysia)

INIS

Country of Publication
Malaysia
Country of Input or Organization
Malaysia
INIS RN
41031049
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; CALCULATION METHODS; GAMMA RADIATION; GAMMA SPECTROSCOPY; PEAKS
Descriptors DEC
ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; RADIATIONS; SPECTROSCOPY