Published January 2011
| Version v1
Journal article
Development of a plastic scintillator based large area ground surface contamination monitor
- 1. Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai (India)
Description
This paper presents the features of a recently developed ground surface contamination monitor based on wide area plastic scintillation detector and investigates the design and performance characteristics of the system. The system incorporates an innovative methodology for suppressing the general background radiation influence on the detector response. The methodology involves the incorporation of a suitably positioned gamma sensitive Geiger Muller tube, the output of which is used to compensate for the general background interference in the response from the plastic scintillation detector. This paper reports on the performance evaluation of the system and the parameters considered for setting the alarm levels. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection and Environment
- Journal Volume
- 34
- Journal Issue
- 1
- Journal Page Range
- p. 41-43
- CODEN
- RPREFM
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 43082204
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- GEIGER-MUELLER COUNTERS; PERFORMANCE TESTING; PLASTIC SCINTILLATION DETECTORS; RADIATION PROTECTION; SPECIFICATIONS
- Descriptors DEC
- MEASURING INSTRUMENTS; RADIATION DETECTORS; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; TESTING
Optional Information
- Notes
- 5 refs., 2 figs.