Published 1981 | Version v1
Journal article

Radiation grafting of N-vinylpyrrolidone into silicone tubes. Synthesis of polymers with improved hemocompatibility and implantation tests in lambs

  • 1. Centre National de la Recherche Scientifique, 94 -Thiais (France). Lab. de Chimie Macromoleculaire sous Rayonnement
  • 2. Compiegne Univ., 60 (France). Lab. de Technologie Enzymatique

Description

NVP was grafted onto silicone 2 mm I.D. tubes by the 'direct' radiation method. Homogeneous bulk grafting was carried out in toluene solution; the best results were obtained in 70% monomer solution, so it was chosen for preparing the samples for the implantations (in vivo tests). In 7 days implantation test the significant improvement of blood compatibility corresponds to grafting ratios higher than 40%. Surface limited grafting is realized in dilute aqueous solutions of NVP. The thickness of the grafted layer inside and outside the tubes is measured after coloration and observation with a microscope. The surface grafting ratio is determined by multiple reflection I.R. analysis. (author)

Additional details

Publishing Information

Journal Title
Radiat. Phys. Chem.
Journal Volume
18
Journal Issue
5-6
Series
Radiat. Phys. Chem.
Journal Page Range
1203-1206
ISSN
0146-5724

Conference

Title
Trends in radiation processing. Transactions of the 3. international meeting on radiation processing.
Dates
26 - 30 Oct 1980.
Place
Tokyo, Japan.

Optional Information

Notes
Vol. 3. Contributed papers edited by L. Silverman.