Published January 1997 | Version v1
Journal article

Radiation-field mapping of insect irradiation canisters

  • 1. National Inst. of Standards and Technology - PL, Gaithersburg, MD (United States). Ionizing Radiation Div.
  • 2. United States Department of Agriculture, Riverdale, MD (United States). Animal and Plant Health Inspection Service

Description

Dosimetry methods developed at NIST for mapping ionizing radiation fields were applied to canisters used in 137Cs dry-source irradiators designed for insect sterilization. The method of mapping the radiation fields inside of these canisters as they cycled through the gamma-ray irradiators involved the use of radiochromic films, which increase in optical density proportionately to the absorbed dose. A dosimeter film array in a cardboard phantom was designed to simulate the average insect pupae density and to map the dose within the full volume of the canister; the calibrated films were read using a laser scanning densitometer. Previously used dosimetric methods did not allow for the spatial resolution that is possible with these films. Results indicate that this dose-mapping technique is a powerful method of evaluating a variety of radiation fields of commercial radiation sources, with promising applications as a means of dose validation and quality control. (Author)

Additional details

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
48
Journal Issue
1
Journal Page Range
p. 117-125.
ISSN
0969-8043
CODEN
ARISEF