Digital radiography: are the manufacturers' settings too high? Optimisation of the Kodak digital radiography system with aid of the computed radiography dose index

  • 1. UCD School of Diagnostic Imaging, Herbert Avenue, Dublin 4 (Ireland)

Description

Manufacturers offer exposure indices as a safeguard against overexposure in computed radiography, but the basis for recommended values is unclear. This study establishes an optimum exposure index to be used as a guideline for a specific CR system to minimise radiation exposures for computed mobile chest radiography, and compares this with manufacturer guidelines and current practice. An anthropomorphic phantom was employed to establish the minimum milliamperes consistent with acceptable image quality for mobile chest radiography images. This was found to be 2 mAs. Consecutively, 10 patients were exposed with this optimised milliampere value and 10 patients were exposed with the 3.2 mAs routinely used in the department of the study. Image quality was objectively assessed using anatomical criteria. Retrospective analyses of 717 exposure indices recorded over 2 months from mobile chest examinations were performed. The optimised milliampere value provided a significant reduction of the average exposure index from 1840 to 1570 (p<0.0001). This new ''optimum'' exposure index is substantially lower than manufacturer guidelines of 2000 and significantly lower than exposure indices from the retrospective study (1890). Retrospective data showed a significant increase in exposure indices if the examination was performed out of hours. The data provided by this study emphasise the need for clinicians and personnel to consider establishing their own optimum exposure indices for digital investigations rather than simply accepting manufacturers' guidelines. Such an approach, along with regular monitoring of indices, may result in a substantial reduction in patient exposure. (orig.)

Availability note (English)

Available from: http://dx.doi.org/10.1007/s00330-001-1230-0

Additional details

Identifiers

Publishing Information

Journal Title
European Radiology
Journal Volume
12
Journal Issue
9
Journal Page Range
p. 2381-2387
ISSN
0938-7994
CODEN
EURAE3

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
33046789
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CHEST; DIGITAL SYSTEMS; IMAGES; INDEXES; MONITORING; OPTIMIZATION; PATIENTS; PERSONNEL; PHANTOMS; QUALITY CONTROL; RECOMMENDATIONS
Descriptors DEC
BODY; CONTROL; DOCUMENT TYPES; MOCKUP; STRUCTURAL MODELS