Published September 1987 | Version v1
Journal article

Microcomputer-controlled perturbed angular correlation spectrometer

  • 1. Department of Physics, Oregon State University, Corvallis, Oregon 97331

Description

A four-detector time-differential perturbed angular correlation spectrometer controlled by an inexpensive microcomputer is described. The use of twin-single-channel analyzers permits each detector to identify both gamma rays in the cascade. An improved gating logic system reduces the MCA/computer deadtime and allows larger sample activity and faster data accumulation. The twin-single-channel analyzer and gating circuit additions also permit simultaneous accumulation of two statistically independent sets of four-channel spectra. The spectrometer provides on-line display of raw data and the perturbation function derived from either set of four-channel spectra

Additional details

Publishing Information

Journal Title
Rev. Sci. Instrum.
Journal Volume
58
Journal Issue
9
Series
Rev. Sci. Instrum.
Journal Page Range
1694-1698
ISSN
0034-6748
CODEN
RSINA