Published September 1987
| Version v1
Journal article
Microcomputer-controlled perturbed angular correlation spectrometer
Creators
- 1. Department of Physics, Oregon State University, Corvallis, Oregon 97331
Description
A four-detector time-differential perturbed angular correlation spectrometer controlled by an inexpensive microcomputer is described. The use of twin-single-channel analyzers permits each detector to identify both gamma rays in the cascade. An improved gating logic system reduces the MCA/computer deadtime and allows larger sample activity and faster data accumulation. The twin-single-channel analyzer and gating circuit additions also permit simultaneous accumulation of two statistically independent sets of four-channel spectra. The spectrometer provides on-line display of raw data and the perturbation function derived from either set of four-channel spectra
Additional details
Publishing Information
- Journal Title
- Rev. Sci. Instrum.
- Journal Volume
- 58
- Journal Issue
- 9
- Series
- Rev. Sci. Instrum.
- Journal Page Range
- 1694-1698
- ISSN
- 0034-6748
- CODEN
- RSINA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 18090987
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANGULAR CORRELATION; DATA ACQUISITION; DATA PROCESSING; DISTURBANCES; ELECTRONIC CIRCUITS; EQUIPMENT INTERFACES; GAMMA SPECTROMETERS; GAMMA SPECTROSCOPY; MICROPROCESSORS; MULTI-CHANNEL ANALYZERS; ON-LINE CONTROL SYSTEMS; SPECIFICATIONS
- Descriptors DEC
- CONTROL SYSTEMS; CORRELATIONS; ELECTRONIC EQUIPMENT; EQUIPMENT; MEASURING INSTRUMENTS; MICROELECTRONIC CIRCUITS; ON-LINE SYSTEMS; PULSE ANALYZERS; SPECTROMETERS; SPECTROSCOPY