The half-life of the positron emitter 62Cu has been measured using 4πβ liquid scintillation (LS) spectrometry and counting in the National Institute of Standards and Technology (NIST) 4π-γ re-entrant ionization chamber (IC). A total of 12 independent LS measurements and two independent IC measurements were made, all of which spanned two to four half-lives. Weighted nonlinear fits of the data resulted in half-lives of 9.68 min ± 0.04 min and 9.673 min ± 0.026 min (uncertainties are combined standard uncertainties) for the LS and IC data, respectively (a difference of only 0.07%). The weighted mean of these two values is 9.67 min ± 0.03 min, which is shorter than the ENSDF-recommended value of 9.74 min ± 0.02 min by 0.7%. Since this radionuclide is being investigated for use in positron emission tomography (PET), this difference in the half-life has an important effect on the estimation of the amount of radioactivity administered to patients when decay corrections must be applied. (author)

Additional details

Publishing Information

Journal Title
Journal of Physics. G, Nuclear and Particle Physics (Online)
Journal Volume
23
Journal Issue
11
Journal Page Range
p. 1707-1714
ISSN
1361-6471

Optional Information

Notes
18 refs; This record replaces 31044086