Electron beam characteristics at extended source-to-surface distances for irregular cut-outs
- 1. Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Bangalore (India)
Description
Electron beam therapy is widely used in the management of cancers. The rapid dose fall-off and the short range of an electron beam enable the treatment of lesions close to the surface, while sparing the underlying tissues. In an extended source-to-surface (SSD) treatment with irregular field sizes defined by cerrobend cutouts, underdosage of the lateral tissue may occur due to reduced beam flatness and uniformity. To study the changes in the beam characteristics, the depth dose, beam profile, and isodose distributions were measured at different SSDs for regular 10x10 cm2 and 15x15 cm2 cone, and for irregular cutouts of field size 6.5x9 cm2 and 11.5x15 cm2 for beam energies ranging from 6 to 20 MeV. The PDD, beam flatness, symmetry and uniformity index were compared. For lower energy (6 MeV), there was no change in the depth of maximum dose (R100) as SSD increased, but for higher energy (20 MeV), the R100 depth increased from 2 cm to 3 cm as SSD increased. This shows that as SSD increases there is an increase in the depth of the maximum dose for higher energy beams. There is a +7 mm shift in the R100 depth when compared with regular and irregular field sizes. The symmetry was found to be within limits for all the field sizes as the treatment distance extended as per International Electrotechnical Commission (IEC) protocol. There was a loss of beam flatness for irregular fields and it was more pronounced for lower energies as compared with higher energies, so that the clinically useful isodose level (80% and 90%) width decreases with increase in SSD. This suggests that target coverage at extended SSD with irregular cut-outs may be inadequate unless relatively large fields are used. (author)
Additional details
Publishing Information
- Journal Title
- Journal of Medical Physics
- Journal Volume
- 35
- Journal Issue
- 4
- Journal Page Range
- p. 207-214
- CODEN
- JMPHFE
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 42014346
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CADMIUM; CERROBEND ALLOYS; DEPTH DOSE DISTRIBUTIONS; ELECTRON BEAMS; ELECTRON DOSIMETRY; NEOPLASMS; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- ALLOYS; BEAMS; BISMUTH ALLOYS; BISMUTH BASE ALLOYS; CADMIUM ALLOYS; DISEASES; DOSES; DOSIMETRY; ELEMENTS; LEAD ALLOYS; LEPTON BEAMS; MEDICINE; METALS; NUCLEAR MEDICINE; PARTICLE BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; SPATIAL DOSE DISTRIBUTIONS; THERAPY; TIN ALLOYS
Optional Information
- Notes
- 8 refs., 4 figs., 3 tabs.