Automated high-dose rate brachytherapy treatment planning for a single-channel vaginal cylinder applicator

  • 1. Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, TX 75390, United States of America (United States)

Description

High dose rate (HDR) brachytherapy treatment planning is conventionally performed manually and/or with aids of preplanned templates. In general, the standard of care would be elevated by conducting an automated process to improve treatment planning efficiency, eliminate human error, and reduce plan quality variations. Thus, our group is developing AutoBrachy, an automated HDR brachytherapy planning suite of modules used to augment a clinical treatment planning system. This paper describes our proof-of-concept module for vaginal cylinder HDR planning that has been fully developed. After a patient CT scan is acquired, the cylinder applicator is automatically segmented using image-processing techniques. The target CTV is generated based on physician-specified treatment depth and length. Locations of the dose calculation point, apex point and vaginal surface point, as well as the central applicator channel coordinates, and the corresponding dwell positions are determined according to their geometric relationship with the applicator and written to a structure file. Dwell times are computed through iterative quadratic optimization techniques. The planning information is then transferred to the treatment planning system through a DICOM-RT interface. The entire process was tested for nine patients. The AutoBrachy cylindrical applicator module was able to generate treatment plans for these cases with clinical grade quality. Computation times varied between 1 and 3 min on an Intel Xeon CPU E3-1226 v3 processor. All geometric components in the automated treatment plans were generated accurately. The applicator channel tip positions agreed with the manually identified positions with submillimeter deviations and the channel orientations between the plans agreed within less than 1 degree. The automatically generated plans obtained clinically acceptable quality. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6560/aa637e

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
62
Journal Issue
11
Journal Page Range
p. 4361-4374
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
49095198
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
BRACHYTHERAPY; COMPUTERIZED TOMOGRAPHY; DOSE RATES; ERRORS; IMAGE PROCESSING; ITERATIVE METHODS; PATIENTS; PLANNING; THERAPEUTIC DOSES
Descriptors DEC
CALCULATION METHODS; DIAGNOSTIC TECHNIQUES; DOSES; MEDICINE; NUCLEAR MEDICINE; PROCESSING; RADIOLOGY; RADIOTHERAPY; THERAPY; TOMOGRAPHY