Pelvimetry and patient acceptability compared between open 0.5-T and closed 1.5-T MR systems
Creators
- 1. Institute of Diagnostic Radiology, University Hospital Zurich, Raemistrasse 100, 8091 Zurich (Switzerland)
- 2. Department of Obstetrics, University Hospital Zurich, Raemistrasse 100, 8091 Zurich (Switzerland)
- 3. Institute of Psychosocial Medicine, University Hospital Zurich, Raemistrasse 100, 8091 Zurich (Switzerland)
- 4. Department of Biostatistics, University Hospital Zurich, Sumatrastrasse 30, 8006 Zurich (Switzerland)
- 5. Department of Otorhinolaryngology, University Hospital Zurich, Raemistrasse 100, 8091 Zurich (Switzerland)
Description
Our objective was to compare maternal pelvimetry and patient acceptability between open low-field (0.5-T) and closed 1.5-T MR systems. Thirty women referred for pelvimetry (pregnant: n=15) were scanned twice in the supine position, once in the vertical open system and once in the closed system. Each patient completed a comfort and acceptability questionnaire. Pelvimetric and questionnaire data were compared between systems. Total scan time was double in the open system (7:52±1:47 vs 3:12±1:20 min). Poor image quality in the open system prevented assessment of interspinous and intertuberous diameters in one woman and all measurements in another, both pregnant, with abdominal circumferences >120 cm. The open system was much more acceptable in terms of claustrophobia and confinement (both p<0.01). Claustrophobia interrupted one closed examination. Thirty-three percent of pregnant women in both systems reported fear of fetal harm. Sixty percent of all women preferred the open system, 7% the closed system, and 33% had no preference. Limits of agreement of 3-5% from the mean for all diameters confirmed good pelvimetric reproducibility. Women's preference for open-system MR pelvimetry is feasible with abdominal circumferences ≤120 cm. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-002-1400-8Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 12
- Journal Issue
- 12
- Journal Page Range
- p. 2898-2905
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 34011220
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; IMAGES; NMR IMAGING; PATIENTS; PELVIS; WOMEN
- Descriptors DEC
- ANIMALS; BODY; DIAGNOSTIC TECHNIQUES; EVALUATION; FEMALES; MAMMALS; MAN; PRIMATES; VERTEBRATES