Pelvimetry and patient acceptability compared between open 0.5-T and closed 1.5-T MR systems

  • 1. Institute of Diagnostic Radiology, University Hospital Zurich, Raemistrasse 100, 8091 Zurich (Switzerland)
  • 2. Department of Obstetrics, University Hospital Zurich, Raemistrasse 100, 8091 Zurich (Switzerland)
  • 3. Institute of Psychosocial Medicine, University Hospital Zurich, Raemistrasse 100, 8091 Zurich (Switzerland)
  • 4. Department of Biostatistics, University Hospital Zurich, Sumatrastrasse 30, 8006 Zurich (Switzerland)
  • 5. Department of Otorhinolaryngology, University Hospital Zurich, Raemistrasse 100, 8091 Zurich (Switzerland)

Description

Our objective was to compare maternal pelvimetry and patient acceptability between open low-field (0.5-T) and closed 1.5-T MR systems. Thirty women referred for pelvimetry (pregnant: n=15) were scanned twice in the supine position, once in the vertical open system and once in the closed system. Each patient completed a comfort and acceptability questionnaire. Pelvimetric and questionnaire data were compared between systems. Total scan time was double in the open system (7:52±1:47 vs 3:12±1:20 min). Poor image quality in the open system prevented assessment of interspinous and intertuberous diameters in one woman and all measurements in another, both pregnant, with abdominal circumferences >120 cm. The open system was much more acceptable in terms of claustrophobia and confinement (both p<0.01). Claustrophobia interrupted one closed examination. Thirty-three percent of pregnant women in both systems reported fear of fetal harm. Sixty percent of all women preferred the open system, 7% the closed system, and 33% had no preference. Limits of agreement of 3-5% from the mean for all diameters confirmed good pelvimetric reproducibility. Women's preference for open-system MR pelvimetry is feasible with abdominal circumferences ≤120 cm. (orig.)

Availability note (English)

Available from: http://dx.doi.org/10.1007/s00330-002-1400-8

Additional details

Identifiers

Publishing Information

Journal Title
European Radiology
Journal Volume
12
Journal Issue
12
Journal Page Range
p. 2898-2905
ISSN
0938-7994
CODEN
EURAE3

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
34011220
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COMPARATIVE EVALUATIONS; IMAGES; NMR IMAGING; PATIENTS; PELVIS; WOMEN
Descriptors DEC
ANIMALS; BODY; DIAGNOSTIC TECHNIQUES; EVALUATION; FEMALES; MAMMALS; MAN; PRIMATES; VERTEBRATES