Published February 1, 2002 | Version v1
Journal article

Operation of MSGC+GEM detectors in a high rate environment

Description

The assembly and test of 18 MSGC+GEM modules each equipped with four wedge-shaped substrates is described. Measurements of gain and current were performed to determine the fields for maximum transparency of the GEM. These detectors were operated for five weeks under a high rate environment at the Paul Scherrer Institute (Switzerland) to study their stability and uniformity. The large scale MSGC+GEM system was operated for 376 h at the nominal working point of signal to noise around 37 with a negligible sparking rate. A safety margin of a factor of two to three in signal to noise guarantees a safe operation under LHC like conditions

Additional details

Identifiers

PII
S016890020101765X;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
478
Journal Issue
1-2
Journal Page Range
p. 250-253
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34001276
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CHARGED PARTICLE DETECTION; ELECTRON MULTIPLIER DETECTORS; GAIN; SIGNAL-TO-NOISE RATIO; STABILITY
Descriptors DEC
AMPLIFICATION; DETECTION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS

Optional Information

Copyright