Published October 1, 2004 | Version v1
Journal article

Evaluation and Monte Carlo modelling of the response function of the Leake neutron area survey instrument

Description

All available measured data for the response characteristics of the Leake counter have been gathered together. These data, augmented by previously unpublished work, have been compared to Monte Carlo simulations of the instrument's response characteristics in the energy range from thermal to 20 MeV. A response function has been derived, which is recommended as the best currently available for the instrument. Folding this function with workplace energy distributions has enabled an assessment of the impact of this new response function to be made. Similar work, which will be published separately, has been carried out for the NM2 and the Studsvik 2202D neutron area survey instruments

Additional details

Identifiers

DOI
10.1016/j.nima.2004.05.086;
PII
S0168900204011076;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
531
Journal Issue
3
Journal Page Range
p. 596-606
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2004 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.