Air Kerma Calibration in Mammography of Thermoluminescence and Semiconductor Dosemeters against an Ionisation Chamber
Creators
Description
The ionisation chamber is often used as the reference dosemeter for dose measurements in radiology. Nevertheless it might be convenient for practical reasons to use thermoluminescence or semiconductor dosemeters as measurement dosemeters. A comparative study is presented using the main dosimetric properties of thermoluminescence and semiconductor dosemeters in mammographic X ray beams (molybdenum anode). The energy and air kerma responses are presented as well as the calibration procedure allowing transfer from the reference dosemeter to the measurement dosemeters. The ionisation chamber used was a plane parallel chamber model 96035B (Keithley, USA). It has been calibrated in terms of air kerma free-in-air at Nederland MeetInstitut for tube voltages between 20 and 35 kV in un attenuated beams. Thermoluminescence dosemeters are GR200A (material, LiF:Mg,Cu,P) and the semi-conductor is PMX3 (sensor ADI/8) and R100 detector from RTI. For both dosemeters, the energy response has been studied and entrance skin air kerma (ESAK) and entrance skin dose (ESD) performed with PMMA plates in order to validate the internal calibration procedure. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 80
- Journal Issue
- 1-3
- Journal Page Range
- p. 239-241
- ISSN
- 0144-8420
- CODEN
- RPDODE
Conference
- Title
- Workshop on reference doses and quality in medical imaging
- Dates
- 23-25 Oct 1997
- Place
- Luxembourg (Luxembourg)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- Kazakhstan
- INIS RN
- 31044181
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPARATIVE EVALUATIONS; DOSEMETERS; KERMA; RADIOLOGY; SEMICONDUCTOR DETECTORS; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- DOSEMETERS; EVALUATION; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTORS