Published June 1990
| Version v1
Journal article
Application of a charge-division, position-sensitive proportional counter to high-resolution PIXE measurements
- 1. Hosei Univ., Koganei, Tokyo (Japan). Research Center of Ion Beam Technology
Description
A charge-division, position-sensitive proportional counter with a carbon-fiber anode and its associated signal processor has been developed. This counter is very small in volume with a simple structure and vacuum proof. The spatial resolution of 0.36 mm was achieved under a gas pressure of 1.7x105 Pa. A correlation factor between the position and the channel number was measured to be 0.97, indicating good linearity of the system. The detector was combined with a crystal spectrometer in order to measure spectra of particle-induced characteristic X rays. Energy resolutions measured at Si Kα (1.74 keV) and Fe Kα (6.4 keV) X rays are 4 and 34 eV, respectively. (author)
Additional details
Publishing Information
- Journal Title
- International Journal of PIXE
- Journal Volume
- 1
- Journal Issue
- 2
- Journal Page Range
- p. 141-146.
- ISSN
- 0129-0835
- CODEN
- IJPXET
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 26050773
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANODES; CARBON FIBERS; DATA PROCESSING; ENERGY RESOLUTION; PIXE ANALYSIS; POSITRON DETECTION; PROPORTIONAL COUNTERS; X-RAY SPECTROMETERS
- Descriptors DEC
- CHARGED PARTICLE DETECTION; CHEMICAL ANALYSIS; DETECTION; ELECTRODES; FIBERS; MEASURING INSTRUMENTS; NONDESTRUCTIVE ANALYSIS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; SPECTROMETERS; X-RAY EMISSION ANALYSIS