Published November 21, 2013 | Version v1
Journal article

Automatic in vivo portal dosimetry of all treatments

  • 1. Department of Radiation Oncology, The Netherlands Cancer Institute–Antoni van Leeuwenhoek Hospital, Plesmanlaan 121, 1066-CX Amsterdam (Netherlands)

Description

At our institution EPID (electronic portal imaging device) dosimetry is routinely applied to perform in vivo dose verification of all patient treatments with curative intent since January 2008. The major impediment of the method has been the amount of work required to produce and inspect the in vivo dosimetry reports (a time-consuming and labor-intensive process). In this paper we present an overview of the actions performed to implement an automated in vivo dosimetry solution clinically. We reimplemented the EPID dosimetry software and modified the acquisition software. Furthermore, we introduced new tools to periodically inspect the record-and-verify database and automatically run the EPID dosimetry software when needed. In 2012, 95% of our 3839 treatments scheduled for in vivo dosimetry were analyzed automatically (27 633 portal images of intensity-modulated radiotherapy (IMRT) fields, 5551 portal image data of VMAT arcs, and 2003 portal images of non-IMRT fields). The in vivo dosimetry verification results are available a few minutes after delivery and alerts are immediately raised when deviations outside tolerance levels are detected. After the clinical introduction of this automated solution, inspection of the detected deviations is the only remaining work. These newly developed tools are a major step forward towards full integration of in vivo EPID dosimetry in radiation oncology practice. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/58/22/8253

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
58
Journal Issue
22
Journal Page Range
p. 8253-8264
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
47007202
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; COMPUTER CODES; DOSIMETRY; IN VIVO; INSPECTION; PATIENTS; PERSONNEL; RADIATION DOSES; RADIOTHERAPY; SCHEDULES; VERIFICATION
Descriptors DEC
DIAGNOSTIC TECHNIQUES; DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY