3D Dose Verification Using Tomotherapy CT Detector Array
Creators
- 1. Department of Radiation Oncology, University of Virginia, Charlottesville, VA (United States)
- 2. TomoTherapy, Inc., Madison, WI (United States)
Description
Purpose: To evaluate a three-dimensional dose verification method based on the exit dose using the onboard detector of tomotherapy. Methods and Materials: The study included 347 treatment fractions from 24 patients, including 10 prostate, 5 head and neck (HN), and 9 spinal stereotactic body radiation therapy (SBRT) cases. Detector sonograms were retrieved and back-projected to calculate entrance fluence, which was then forward-projected on the CT images to calculate the verification dose, which was compared with ion chamber and film measurement in the QA plans and with the planning dose in patient plans. Results: Root mean square (RMS) errors of 2.0%, 2.2%, and 2.0% were observed comparing the dose verification (DV) and the ion chamber measured point dose in the phantom plans for HN, prostate, and spinal SBRT patients, respectively. When cumulative dose in the entire treatment is considered, for HN patients, the error of the mean dose to the planning target volume (PTV) varied from 1.47% to 5.62% with a RMS error of 3.55%. For prostate patients, the error of the mean dose to the prostate target volume varied from –5.11% to 3.29%, with a RMS error of 2.49%. The RMS error of maximum doses to the bladder and the rectum were 2.34% (–4.17% to 2.61%) and 2.64% (–4.54% to 3.94%), respectively. For the nine spinal SBRT patients, the RMS error of the minimum dose to the PTV was 2.43% (–5.39% to 2.48%). The RMS error of maximum dose to the spinal cord was 1.05% (–2.86% to 0.89%). Conclusions: An excellent agreement was observed between the measurement and the verification dose. In the patient treatments, the agreement in doses to the majority of PTVs and organs at risk is within 5% for the cumulative treatment course doses. The dosimetric error strongly depends on the error in multileaf collimator leaf opening time with a sensitivity correlating to the gantry rotation period.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ijrobp.2010.12.043Additional details
Identifiers
- DOI
- 10.1016/j.ijrobp.2010.12.043;
- PII
- S0360-3016(11)00031-9;
Publishing Information
- Journal Title
- International Journal of Radiation Oncology, Biology and Physics
- Journal Volume
- 82
- Journal Issue
- 2
- Journal Page Range
- p. 1013-1020
- ISSN
- 0360-3016
- CODEN
- IOBPD3
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44016464
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BLADDER; CAT SCANNING; CT-GUIDED RADIOTHERAPY; DOSIMETRY; HAZARDS; HEAD; IONIZATION CHAMBERS; NECK; PATIENTS; PHANTOMS; PLANNING; PROSTATE; RADIATION DOSES; RECTUM; SENSITIVITY; SPINAL CORD; VERIFICATION
- Descriptors DEC
- BODY; CENTRAL NERVOUS SYSTEM; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DIGESTIVE SYSTEM; DOSES; GASTROINTESTINAL TRACT; GLANDS; INTESTINES; LARGE INTESTINE; MALE GENITALS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NERVOUS SYSTEM; NUCLEAR MEDICINE; ORGANS; RADIATION DETECTORS; RADIOLOGY; RADIOTHERAPY; STRUCTURAL MODELS; THERAPY; TOMOGRAPHY; URINARY TRACT
Optional Information