Energy-correction photon counting pixel for photon energy extraction under pulse pile-up
Description
A photon counting detector (PCD) has been proposed as an alternative solution to an energy-integrating detector (EID) in medical imaging field due to its high resolution, high efficiency, and low noise. The PCD has expanded to variety of fields such as spectral CT, k-edge imaging, and material decomposition owing to its capability to count and measure the number and the energy of an incident photon, respectively. Nonetheless, pulse pile-up, which is a superimposition of pulses at the output of a charge sensitive amplifier (CSA) in each PC pixel, occurs frequently as the X-ray flux increases due to the finite pulse processing time (PPT) in CSAs. Pulse pile-up induces not only a count loss but also distortion in the measured X-ray spectrum from each PC pixel and thus it is a main constraint on the use of PCDs in high flux X-ray applications. To minimize these effects, an energy-correction PC (ECPC) pixel is proposed to resolve pulse pile-up without cutting off the PPT by adding an energy correction logic (ECL) via a cross detection method (CDM). The ECPC pixel with a size of 200×200 µm2 was fabricated by using a 6-metal 1-poly 0.18 µm CMOS process with a static power consumption of 7.2 μW/pixel. The maximum count rate of the ECPC pixel was extended by approximately three times higher than that of a conventional PC pixel with a PPT of 500 nsec. The X-ray spectrum of 90 kVp, filtered by 3 mm Al filter, was measured as the X-ray current was increased using the CdTe and the ECPC pixel. As a result, the ECPC pixel dramatically reduced the energy spectrum distortion at 2 Mphotons/pixel/s when compared to that of the ERCP pixel with the same 500 nsec PPT.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2017.02.084Additional details
Identifiers
- DOI
- 10.1016/j.nima.2017.02.084;
- PII
- S0168-9002(17)30307-8;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 856
- Journal Page Range
- p. 36-46
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48085884
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- AMPLIFIERS; CADMIUM TELLURIDES; COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; CORRECTIONS; ENERGY SPECTRA; FILTERS; PHOTON COUNTING; PHOTONS; PULSES; RADIATION FLUX; RESOLUTION; SEMICONDUCTOR DEVICES; X RADIATION; X-RAY SPECTRA
- Descriptors DEC
- BOSONS; CADMIUM COMPOUNDS; CHALCOGENIDES; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; ELECTRONIC EQUIPMENT; ELEMENTARY PARTICLES; EQUIPMENT; EVALUATION; IONIZING RADIATIONS; MASSLESS PARTICLES; RADIATIONS; SPECTRA; TELLURIDES; TELLURIUM COMPOUNDS; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2017 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.