Response of TLD-100TM microtubes to two RQR3 quality radiation beams

  • 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)
  • 2. Universidad Nacional Autonoma de Mexico (Mexico)

Description

The present work compares the response of TLD-100TM microcubes to two RQR 3 diagnostic radiology reference quality radiation beams, defined by IEC-61267 norm, aiming to evaluate the detectability of TLD-100TM energy dependence reported in literature within the same reference quality radiation range. TLD-100TM microcubes reproducibility is assessed through the response of a second set of TLD-100TM microcubes, evaluated in a second thermoluminescence reader, to the RQR 3 diagnostic radiology reference quality radiation beam implemented at the Laboratorio de Calibracao de Instrumentos of IPEN, Sao Paulo, SP, Brazil. The dependence of TLD-100TM microcubes TL response was not detectable in these conditions and the reproducibility of the measurements is 90,2%. (author)